866.4.Maxtec www.maxtecinc.com12
6.0 CHANGING THE OXYGEN SENSOR
Turn off analyzer. Disconnect sensor from
sensor cable.
Remove sensor from
sensor port. Dispose of
sensor according to local
guidelines.
1
2
3
7.0 CLEANING AND MAINTENANCE
Store the MaxVenturi at a temperature similar to that of its daily use.
Device Cleaning:
When cleaning or disinfecting the exterior of the MaxVenturi, take appropriate care
to prevent any solution from entering the device. Do not immerse device in fluids.
MaxVenturi external surfaces may be cleaned using a cloth moistened with 65-70%
isopropyl alcohol/water solution or a germicidal wipe.
Oxygen Sensor Cleaning:
• Clean the sensor with a cloth moistened with 65-70% isopropyl alcohol/water
solution.
• Maxtec does not recommend use of spray disinfectants because they can contain
salts which can accumulate in the sensor membrane and impair readings.
• The oxygen sensor is not intended for steam, ethylene oxide or radiation
sterilization.
Note: For new sensor installation, follow procedure for new sensor installation in
section 2.1.