866.4.Maxtec www.maxtecinc.com4
2.2 Calibration
A new calibration is required when:
• The measured O
2
percentage in 100% O
2
is below 97.0%.
• The measured O
2
percentage in 100% O
2
is above 103.0%.
• The CAL reminder icon is blinking at the bottom of the LCD.
• You are unsure about the displayed O
2
percentage, see Factors Influencing
Accurate Readings in section 3.0.
The MaxVenturi can be calibrated at 100% oxygen or room oxygen (20.9%). The One
Touch calibration will assume one of these two concentrations.
2.2.1 Room Air Calibration
Caution: Turn off all flow of gas to the MaxVenturi before calibration in room air.
Calibrating the oxygen sensor in a gas concentration other than room air (20.9%) will
result in an incorrect measurement of oxygen concentration.
1
2
3
...2 minutes
5
Shut off Oxygen supply. Remove sensor from
port.
Wait two (2) minutes
for sensor to equilibrate
in room air.
Press and hold CAL
button for three (3)
seconds. Wait for
display to
read "20.9%".
Place sensor in port.