3.0 FACTORS INFLUENCING ACCURATE READINGS
3.1 Elevation/Pressure Changes
• Changes in elevation result in a reading error of approximately 1% of reading per
250 feet.
• A change in altitude greater than 500 ft will require sensor recalibration.
• This device does not automatically compensate for changes in barometric pressure
or altitude. If the device is moved to a location of a different altitude, it must be
recalibrated before use (see section 2.2).
3.2 Temperature Effects
The MaxVenturi will read correctly (within ±3%) when operating at thermal equilibrium
within the operating temperature range of 15˚C - 40˚C (59˚F - 104˚F). The device
must be thermally stable when calibrated and allowed to thermally stabilize after
experiencing temperature changes before readings are accurate.
For these reasons it is recommended to:
• Perform the calibration procedure at a temperature close to the temperature at
which the device will be operated.
• Allow adequate time for the sensor to equilibrate to a new ambient temperature.
ATTENTION: “CAL Err St” may result from a sensor that has not reached
thermal equilibrium.
4.0 CALIBRATION ERRORS AND ERROR CODES
The analyzer has a self test feature built into the software to detect faulty calibrations,
oxygen sensor failures, and low battery voltage. The tests and actions to take if an error
occurs are:
E02: NO SENSOR ATTACHED
• Disconnect and reconnect sensor. Unit should perform an auto calibration
and should read 20.9%. If not, contact Customer Service for possible sensor
replacement.
E03: NO VALID CALIBRATION DATA AVAILABLE
• Make sure unit has reached thermal equilibrium.
• Perform calibration as described in this manual.
E04: BATTERY BELOW MINIMUM OPERATING VOLTAGE
• Replace batteries.
CAL Err St: O
2
SENSOR READING NOT STABLE
• Wait for displayed oxygen reading to stabilize when calibrating the device at
100% oxygen.
• Wait for unit to reach thermal equilibrium. Note: this can take up to one half
hour if the device is stored in temperatures outside the specified operating
temperature range.
CAL Err lo: SENSOR VOLTAGE TOO LOW
• Repeat calibration routine as described in this manual. If unit repeats this error
more than three times, contact Customer Service.